How to use
For Smart MicroVision NX. Structured cases go through Path-Snap, not personal messaging.
Focus a well-lit field through the eyepieces before attaching the phone. Start at 10×. Keep room light off the eyepiece-camera junction. Flash off.
Fit NX
- Focus the slide.
- Place the phone or tablet on the platform.
- Measure from the lower edge to the centre of the main 1× camera.
- Set the red line or arrow to that distance.
- Seat the adaptor and align the camera with the eyepiece using the lit disc on screen.
- Tilt until the field is circular and even.
Record phone model, scale, and orientation. Capture an overview and a high-power field. Do not send near-duplicates.
Phone or tablet
Prefer a dedicated laboratory device. Silent / Do Not Disturb. Main 1× camera. Flash, HDR, auto-rotate, GPS, and cloud upload off unless policy requires them. Lock focus and exposure when sharp. Keep battery above half and the screen awake.
Path-Snap
Use Path-Snap only for specimen types listed on its page. Teaching sets without Path-Snap: strip identifiers, one patient, low- and high-power views, a stated question. No faces or LIS screens.